Interventional Cardiovascular Magnetic Resonance Imaging: A New Opportunity for Image-Guided Interventions FREE

Christina E. Saikus, BS; Robert J. Lederman, MD
[+] Author Information

Reprint requests and correspondence: Dr. Robert J. Lederman, Translational Medicine Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Building 10, Room 2c713, MSC1538, Bethesda, Maryland 20892-1538

American College of Cardiology Foundation

J Am Coll Cardiol Img. 2009;2(11):1321-1331. doi:10.1016/j.jcmg.2009.09.002
Published online

  Cardiovascular magnetic resonance (CMR) combines excellent soft-tissue contrast, multiplanar views, and dynamic imaging of cardiac function without ionizing radiation exposure. Interventional cardiovascular magnetic resonance (iCMR) leverages these features to enhance conventional interventional procedures or to enable novel ones. Although still awaiting clinical deployment, this young field has tremendous potential. We survey promising clinical applications for iCMR. Next, we discuss the technologies that allow CMR-guided interventions and, finally, what still needs to be done to bring them to the clinic.

Figures in this Article
3D

3-dimensional

CMR

cardiovascular magnetic resonance

EP

electrophysiology

iCMR

interventional cardiovascular magnetic resonance

RF

radiofrequency

VSD

ventricular septal defect

XFM

X-ray–fused with magnetic resonance imaging

Minimally invasive and catheter-based therapies are targeting increasingly complex pathologies. This agenda requires better procedural image guidance (Table 1). Interventional cardiovascular magnetic resonance (iCMR) proposes to use the superb tissue imaging afforded by cardiovascular magnetic resonance (CMR) as a surrogate for direct surgical visualization. By providing multiple views and real-time functional imaging without ionizing radiation exposure, CMR could guide traditional interventional procedures and enable novel ones. We provide a brief overview of cardiovascular procedures that could benefit from real-time CMR guidance, followed by technical advances helping to move toward clinical testing.

Table Grahic Jump Location
Table 1Comparison of Imaging Modalities for Real-Time Procedural Guidance
Table Footer NoteNote the frame rate and spatial resolution given for CMR are arbitrary and represent a typical compromise between spatial and temporal resolution at 1.5-T using parallel imaging with an acceleration factor of 2.
Catheter-based endovascular interventions

CMR visualizes vessel wall and lumen without exogenous contrast. Interest in atherosclerosis imaging fueled much of the early work in intravascular CMR. Building on internal receiver coils for high-resolution intracardiac 31P nuclear magnetic resonance (1), local endovascular coils were developed to characterize arterial wall composition with improved signal-to-noise ratio and resolution (23). Flow and motion, however, remain significant challenges to high-resolution magnetic resonance (MR) vessel images with realistic scan times in vivo.

Vascular interventions such as angioplasty and stenting under CMR could reduce radiation and iodinated contrast exposure compared with X-ray procedures. The use of CMR may also allow image-based monitoring for serious complications such as vessel wall dissection or perforation (45). Newer stent alloys (including Nitinol and nickel-cobalt-chromium) are more compatible with CMR than older stainless-steel materials. Despite a dearth of other commercial devices suitable for interventional CMR, numerous endovascular procedures have been demonstrated in pre-clinical and early clinical studies. Although coronary catheterization and stenting have been performed under CMR guidance in healthy animals (68), we caution that clinical CMR lacks sufficient spatial or temporal resolution to guide meaningful clinical coronary interventional procedures.

Larger peripheral vessels such as aorta and iliac arteries are more attractive targets for CMR-guided procedures. Simple iliac angioplasty and stenting was performed in human feasibility studies by Manke et al. (9), albeit with limited imaging guidance. Renal artery stenting has been conducted solely under CMR in an animal renal artery stenosis (10). CMR-directed tube endografts repaired porcine abdominal aortic aneurysms (11) and thoracic aortic dissection after distinguishing true and false lumens (12). Aortic coarctation stenting with CMR guidance also has been explored (5). The authors of a clinical pilot study (13) successfully demonstrated CMR-assisted angioplasty for aortic coarctation in combination with X-ray.

Completely blocked vessel segments (chronic total occlusions) are virtually invisible in X-ray angiography where contrast cannot enter the obstructed lumen. The use of CMR can help to depict the mural contour of the occluded vessel and can help avoid perforation. Recanalization of an animal model carotid chronic total occlusion with the use of a custom active CMR device was significantly more successful than X-ray-guided alternatives (14). Newer devices are being developed that clearly indicate the device position during recanalization (15).

Moving beyond the vessels: extra-anatomic bypass, structural heart, and biological therapies

The use of CMR could allow catheter-based procedures to escape the traditional confines of vascular lumens. Extra-anatomic bypass, directly connecting two otherwise-unconnected vascular structures, might be enabled without surgery because of CMR guidance. In transjugular intrahepatic portosystemic shunt procedures, CMR helped reduce the number of needle passes needed to connect the hepatic to portal vein within the liver (16). Mesocaval puncture is an alternative portocaval trajectory enabled by CMR and custom-designed needle devices (17). A nonsurgical Blalock-Taussig subclavian-pulmonary shunt is a tantalizing potential application of this technology.

Structural heart disease interventions may benefit from the use of CMR in visualizing critical anatomy. CMR has guided-needle (18) and laser (19) atrial septal puncture and delivery of Nitinol closure devices for atrial septal defects (2022) in swine. The use of CMR has enhanced positioning (with regard to myocardial fibrous skeleton and coronary arteries) and monitoring of transcatheter (23) and surgical transapical (24) (Figure 1) aortic valve implants in swine with immediate evaluation of valve performance and flow. Hybrid imaging guidance for valve placement is also under development with the use of an adjacent “CMR-friendly” X-ray fluoroscope at the edge of the magnet (25).

Grahic Jump Location
Figure 1

CMR-Guided Transapical Aortic Valve Replacement

(A) Bioprosthesis mounted on a platinum iridium stent with a stainless-steel marker welded on the side of the stent between the commissures. (B) The marker is visible as a dark signal in the cardiovascular magnetic resonance (CMR) and indicates the orientation of the prosthesis. Short-axis view and long-axis view of the implanted prosthesis in a pig under real-time CMR are shown in C and D, respectively. Blue dots are digital markers indicating the coronary ostia whereas the yellow dot shows aortic annulus location. (E) Three-dimensional rendering snapshots show multiple image planes displayed at their relative 3-dimensional position. Images courtesy of Ming Li, PhD, and Keith A. Horvath, MD, Cardiothoracic Surgery Research Program, National Heart, Lung, and Blood Institute.

With the development of novel biologic treatments, including local small molecule, gene, or cellular agents, CMR could play an important role in testing, delivering, and monitoring these therapies. The use of CMR provides enhanced catheter-based endomyocardial targeting and interactive imaging of local cell or drug accumulation and dispersion compared with X-ray or electroanatomic imaging (2627). Several groups (2831) have demonstrated successful delivery or tracking of cell and gene products to myocardial targets in animals by using CMR.

Pediatrics

Pediatric patients stand to gain much from CMR-guided interventional procedures. Children with congenital heart disease often require multiple catheterizations to assess physiological parameters and for treatment that might avoid or delay open-heart surgery. Each interventional procedure, however, is offset by the risk for mechanical complications and possible long-term consequences from X-ray exposure. Children are especially susceptible to radiation injury, and pediatric interventions often are protracted, contributing further to the risk of chromosomal damage and malignancy (32). For medical staff, occupational X-ray exposure risks cancer and cataracts (33), whereas protective lead garments risk chronic orthopedic injury (34).

Razavi et al. (35) brought real-time CMR into the clinic. They guided diagnostic catheterizations and electrophysiological procedures in children and adults with congenital heart disease by using a combined MR/X-ray suite. The use of CMR-guided catheterization has also enhanced the assessment of total or split pulmonary vascular resistance in patients (3637). Beyond invasive diagnostics, CMR might enhance treatment of structural pathology in children and adults with congenital heart disease. Interventions for structural abnormalities such as ventricular septal defect could benefit from improved visualization of the septal defect and close procedural monitoring. Pre-clinical CMR-guided pulmonary artery and valve stenting (38) suggests percutaneous pulmonary valve replacement (39) might be feasible under CMR. As mentioned previously, CMR may enable percutaneous creation and closure of pulmonary-systemic shunts.

Electrophysiology (EP)

Interventional CMR proposes one-stop visualization of cardiac surfaces being mapped, monitoring of ablation, and assessment of ablation lesions. Compared with ablation under direct surgical exposure, conventional catheter ablations of atrial fibrillation or ventricular-tachycardia are lengthy and complex, in part because of inadequate targeting and visualization of ablation lesions. The use of CMR offers value beyond creating baseline roadmaps now widely imported into electroanatomic mapping systems (40). Investigators have begun to correlate ablation lesions to successful rhythm therapy (4142). MR-compatible cardiac monitoring and ablation systems have enabled simple CMR-guided EP procedures (4,4345). (Figure 2) displays a sample CMR-guided EP mapping and recording study (45). Recently, Nazarian et al. (46) at Johns Hopkins demonstrated the feasibility of real-time CMR-guided EP mapping in 2 patients.

Grahic Jump Location
Figure 2

Electrophysiological Cardiac Mapping Under Real-Time CMR Guidance

(Top) Roadmap-based real-time 3-dimensional visualization of the catheter position during recording (red dot) on the magnetic resonance-electrophysiological workstation. The yellow dots in the 3-dimensional rendering of the heart indicate previous mapping positions. (Bottom) In-bore electrophysiological recordings at 2 selected positions showing an atrial signal (left) and a ventricular signal (right). Graphic courtesy of Steffen Weiss, PhD, Philips Healthcare.

Multimodality image guidance

X-ray combined with magnetic resonance imaging, referred to as coregistration or XFM (X-ray fused with magnetic resonance imaging), overlays previously acquired MR images to enhance otherwise-difficult X-ray fluoroscopy procedures. These XFM procedures can benefit from the wide armamentarium of X-ray-compatible catheter equipment, along with additional 3-dimensional (3D) information about target vascular structures afforded by previous CMR. This multimodality approach has been used clinically in complex X-ray procedures such as biopsy of the myocardial free wall (47). The use of XFM also enabled precise endomyocardial injections (48) and a complex repair of membranous ventricular septal defects with significantly decreased fluoroscopy times in animal studies (49). Clinical EP studies already combine CMR roadmaps in commercial electroanatomic mapping systems (40).

In XFM, 3D contours of cardiovascular structures from CMR are overlaid on live X-ray (Figure 3). Views update as the C-arm or table position changes. X-ray and CMR can be coregistered by the use of a shared table space (50), but patient movement causes misregistration. External fiducial marker beads on the patient can align MR and X-ray acquisitions even after patients move (47,51). Static roadmaps also become inaccurate from respiratory and cardiac motion, which can be corrected by the use of more sophisticated techniques (52). Catheter position can be computed from 2 X-ray projections and back-displayed on a 3D MR roadmap.

Grahic Jump Location
Figure 3

X-Ray Fused With MRI

Representative X-ray fused with magnetic resonance imaging (MRI) display with contours from cardiovascular magnetic resonance (CMR) overlaid on live X-ray. Aorta (red), right ventricle (yellow), left endocardial (blue), and epicardial (green) surfaces from CMR shown in 3 dimensions with aortic annulus and ventricular septal defect locations outlined in blue and red, respectively. Multimodality external fiducial markers also are seen numbered.

Interventional MRI in other clinical disciplines

Several other clinical disciplines have embraced interventional MRI to guide critical procedures. Most enjoy targets that are relatively accessible, immobile, or require short, straight-needle devices. Neurosurgeons routinely use pre-procedural MRI with stereotactic frames for biopsies and resections; many conduct intraoperative MRI to guide, for example, tumor resection margins (53) or placement of deep-brain stimulation electrodes (54). Biopsy is also conducted under MRI for suspected liver, breast, and prostate malignancies (55). Tumor cryo- and radiofrequency ablation benefit from MRI guidance, which creates real-time temperature maps of target tissue (56). The use of MRI enhances orthopedic biopsies, injections, and other therapies as well (57). Many of these applications have fueled development of local imaging coils and MR-compatible manipulators and robotics (58).

In this section we review advances in scanner design, imaging, and catheter devices that enable iCMR. We also discuss the remaining hurdles to clinical use.

Interventional suite and scanner selection

Short, wide, closed-bore 1.5- and 3.0-T scanners are now available. These systems provide superior image quality than earlier lower-field open magnets. Wide-bore scanners also allow satisfactory transfemoral access to patients for interventional procedures compared with earlier closed-bore machines. The use of 3.0-T offers the potential to increase resolution and signal-to-noise ratio, but 3.0-T has been challenging to use with workhorse imaging techniques considered essential at 1.5-T.

Most research sites colocalize or combine MRI and X-ray angiography equipment to allow combined procedures and angiographic bailout (Figure 4). Transfer tables or mobile MRI units (59) permit elective intermodality transfer or emergency evacuation. In a properly shielded room, both systems also run independently as MRI scanner and X-ray interventional labs. Fiberoptic or pneumatic headset systems with noise cancellation allow staff and patients to communicate despite loud acoustic MRI noise.

Grahic Jump Location
Figure 4

Interventional CMR Suite

An interventional cardiovascular magnetic resonance (CMR) suite showing adjacent and interoperable magnetic resonance imaging scanner and X-ray angiography labs. A docking gurney allows rapid intermodality transfer. Courtesy of Alexander J. Dick, MD, Sunnybrook Health Sciences Centre, Toronto.

Real-time CMR

Steady-state free precession (often referred to as TrueFISP, FIESTA, or FFE) provides excellent blood and myocardial contrast in electrocardiogram-gated (nonreal-time) acquisitions. Steady-state free precession is now widely available in real time and has become the “workhorse” pulse sequence for interventional procedures. That said, it is no simple feat to acquire, reconstruct via computer, and display CMR images in less than a quarter second. Newer parallel techniques that use large receiver coil arrays and advanced computational techniques (6062) allow even-faster scanning. Coupled with high-performance computing multiple-processor systems, real-time CMR is now available with image updates at 10 frames/s. Other phenomena, such as blood flow and myocardial tags, also can be imaged in real time. Although these MR images have a relatively small pixel matrix compared with X-ray (192 × 128 vs. 1,024 × 1,024), they are comparably rich in information because of extra soft-tissue detail (See Table 1).

Real-time MRI interfaces reconstruct and display images geared for interventional procedure guidance (6364). Multiple concurrent slices can be displayed in 3D to indicate their relative location and orientation (Figure 5A). Where real-time MRI resolution is inadequate, high-resolution previous cine or MR angiography roadmaps can also be incorporated to complement the live imaging ((Figure 5)B and Figure 5C). Real-time user interfaces (Figure 6) are now available from commercial system vendors (65). Foot pedals and related remote controls afford the in-room operator some control over scan planes, but interventional MRI usually requires a dedicated scan operator in addition to the interventionist.

Grahic Jump Location
Figure 5

Real-Time Cardiac MR

Multiplanar real-time cardiac imaging at 4 to 5 frames/s (A) with pre-acquired static cardiac magnetic resonance (MR) image (B) and pulmonary vessel 3-dimensional MR angiography (C) roadmaps.

Grahic Jump Location
Figure 6

Real-Time MRI User Interface Showing MR-Guided Delivery and Biopsy of Encapsulated Human Islet Cells in a Pig

Example of a real-time magnetic resonance imaging (MRI) user interface from Siemens. Similar interfaces are available for Philips and GE systems. Features include scan plane manipulation, independent device channel coloring, 3-dimensional viewing window, static image roadmaps, and quick access to select slice orientations. Figure courtesy of Christine H. Lorenz, PhD, Siemens Corporate Research, Inc., and Aravind Arepally, MD, Johns Hopkins Medical Institutions, Baltimore, Maryland.

When using 2-dimensional imaging techniques, it can be difficult to locate a structure or device that moves out of the selected scan plane. X-ray operators are accustomed to projection views where the entire anatomy or device is visible in the field of view despite its position in the (“through-plane”) third dimension. Independent coloring of multiple active device channels and device-only projection views help overcome this problem (63). Alternatively, computer algorithms can locate and even automatically reposition the scan plane to show the entire active MR device in 3 dimensions (6668). Comparable techniques have been developed to track the position of passive devices (69).

Interventional devices

The same mechanisms that provide unique image contrast during CMR also make most X-ray catheters either invisible or unacceptable for use during iCMR procedures. Metallic cores or braiding that impart catheter pushability and torquability cause imaging artifacts and obscure entire organs. Polymer-only catheters are not visible unless modified and often lack requisite mechanical performance. Conspicuous and safe CMR catheter devices are not generally commercially available but represent an active area of research and development. Interventional CMR devices typically are classified by the mechanism they are visualized in imaging and connection to scanner hardware: passive, semiactive, and active.

Passive devices rely on material properties for visibility during CMR. Material magnetic susceptibility, or how it responds to a magnetic field, can cause inhomogeneities in the main magnetic field that result in signal voids (negative contrast or dark spots) on images. Positive contrast (bright spots) can be generated by the use of paramagnetic T1 shortening contrast agents. Many early passive devices used polymer catheters with ferromagnetic (70) or paramagnetic coatings or rings (7172). Other approaches include filling catheter balloons with CO2 (Figure 7A) (73) or more novel contrast agents such as 19F (74) and hyperpolarized 13C (75) detected with multispectral CMR. Off-resonance imaging techniques (7677) improve the specificity of the device-related signal but usually sacrifice visibility of surrounding anatomy. Recently, several groups (7879) have developed sophisticated nonmetallic guidewires that mimic the mechanical properties of metallic X-ray guidewires. They are conspicuous because of small metal (susceptibility) markers. An example of a passive device visualized through magnetic susceptibility effects is shown in (Figure 8), where a passive guidewire navigates an iliac artery. Passive devices avoid many of the radiofrequency (RF) safety concerns discussed in the “Ensuring iCMR Safety” section and often are simpler to manufacture. However, they remain difficult to discern from background tissue in vivo, especially within curved vascular structures.

Grahic Jump Location
Figure 7

Interventional CMR Devices

Three representative approaches to catheters designed to be tracked using cardiovascular magnetic resonance (CMR). (A) Passive nonbraided balloon catheter filled with CO2 in a right atrium. (B) Active tracking catheter in the right heart. Image courtesy of Michael Bock, PhD, German Cancer Research Center (DKFZ), Heidelberg, Germany. (C) Two-channel active guidewire in the aorta approaching the left heart.

Grahic Jump Location
Figure 8

Iliac Artery Navigation Using a Passive Guidewire

(Top) Polymer guidewire with a Nitinol tip. (Bottom) Magnetic resonance-guided probing of the contralateral iliac artery with the guidewire (arrows) and a catheter. Susceptibility artifacts in cardiovascular magnetic resonance (CMR) images arise from the iron doped guidewire, which is also visible under fluoroscopy. These artifacts partially obscure the target tissue. Images courtesy of Gabriele A. Krombach, MD, Department of Diagnostic Radiology, University Hospital Aachen, Aachen, Germany.

Active devices for interventional CMR incorporate small coils or antennae on independent channels connected to the scanner to track (80) or display the device. “Tracking” requires special CMR pulse sequences to locate the tracking coil in 3D space with the computer-synthesized device position overlaid onto images (Figure 7B). Active device “imaging” or “profiling” allows a device to be depicted uniquely (for example, based on color) on CMR images acquired in real time. An example is the active guidewire in (Figure 7)C. Many of the early coil designs for atheroimaging evolved into active catheters and guidewires devices having loop and loopless (81) antenna configurations. It is challenging to incorporate requisite electrical and mechanical components in appropriately sized designs that simultaneously satisfy operators' functional expectations. Clinical-grade active guidewires and catheters are nearing clinical reality (82), but thorough performance and safety evaluation remains an important step.

Semiactive devices typically contain circuit elements such as inductively coupled markers. These do not require long transmission lines (connecting electrical cables) and need not be connected directly to the scanner but may be more visible than passive devices. Through mutual inductance the signal around the inductively coupled coil can be detected by surface receiver coils. Inductively coupled markers can be incorporated on the end of catheters (8384), but device visualization is then limited to this segment. Further optical tuning (8586) or signal separation (87) techniques may be required to clearly distinguish these devices from background imaging.

Ensuring iCMR safety

MRI uses 3 significant electromagnetic fields. The static, permanent magnetic field, B0, can exert tremendous force and torque on ferromagnetic objects, drawing them into the bore. Consequently, instruments, monitoring equipment, and other implements used near MRI scanners must be nonmagnetic or properly labeled and secured to avoid creating dangerous projectiles in the room. Time-varying magnetic field gradients used to create images may cause peripheral nerve stimulation, which limits the slew rate and therefore imaging speed. The vibration of these gradient coils creates the acoustic noise that can be mitigated by the use of hearing protection and MR-compatible communication systems. Finally, the pulsed electromagnetic RF field, B1, used to energize hydrogen spins during imaging, produces a significant heating risk. Long conductors such as cables connecting local receiver coils, electrocardiogram leads, and active CMR devices, when exposed to the changing magnetic fields, can develop induced currents and heat due to resistive losses. More important, coupling with the electrical field component of B1 can store electrical energy along the device that creates heat near the device tips (88). Several approaches have been explored to minimize active device heating, including detuning devices during transmission when RF energy is highest, and the use of RF chokes (89) or transformers (90) in transmission lines to alter their electrical length. Newer optical transmission lines (91) or wireless coils may provide safer alternatives in the future to relay device MR signals to the scanner.

Pre-clinical demonstrations of iCMR have been enticing. The question remains how to bring this technology into the clinic. The requisite imaging capabilities for iCMR are available on commercial MRI systems. Compatible clinical-grade catheter devices remain largely unavailable absent much commercial investment. Academic centers are now independently developing clinical-grade active and passive catheter tools. Although the pace of development has been slow, future opportunities remain rich.

Kantor  H.L., Briggs  R.W., Balaban  R.S.; In vivo 31P nuclear magnetic resonance measurements in canine heart using a catheter-coil. Circ Res. 55 1984:261-266.
Hurst  G.C., Hua  J., Duerk  J.L., Cohen  A.M.; Intravascular (catheter) NMR receiver probe: preliminary design analysis and application to canine iliofemoral imaging. Magn Reson Med. 24 1992:343-357.
Martin  A.J., Plewes  D.B., Henkelman  R.M.; MR imaging of blood vessels with an intravascular coil. J Magn Reson Imaging. 2 1992:421-429.
Nordbeck  P., Bauer  W.R., Fidler  F.; Feasibility of real-time MRI with a novel carbon catheter for interventional electrophysiology. Circ Arrhythmia Electrophysiol. 2 2009:258-267.
Raval  A.N., Telep  J.D., Guttman  M.A.; Real-time magnetic resonance imaging-guided stenting of aortic coarctation with commercially available catheter devices in swine. Circulation. 112 2005:699-706.
Omary  R.A., Green  J.D., Schirf  B.E., Li  Y., Finn  J.P., Li  D.; Real-time magnetic resonance imaging-guided coronary catheterization in swine. Circulation. 107 2003:2656-2659.
Serfaty  J.M., Yang  X., Foo  T.K., Kumar  A., Derbyshire  A., Atalar  E.; MRI-guided coronary catheterization and PTCA: a feasibility study on a dog model. Magn Reson Med. 49 2003:258-263.
Spuentrup  E., Ruebben  A., Schaeffter  T., Manning  W.J., Gunther  R.W., Buecker  A.; Magnetic resonance-guided coronary artery stent placement in a swine model. Circulation. 105 2002:874-879.
Manke  C., Nitz  W.R., Djavidani  B.; MR imaging-guided stent placement in iliac arterial stenoses: a feasibility study. Radiology. 219 2001:527-534.
Elgort  D.R., Hillenbrand  C.M., Zhang  S.; Image-guided and -monitored renal artery stenting using only MRI. J Magn Reson Imaging. 23 2006:619-627.
Raman  V.K., Karmarkar  P.V., Guttman  M.A.; Real-time magnetic resonance-guided endovascular repair of experimental abdominal aortic aneurysm in swine. J Am Coll Cardiol. 45 2005:2069-2077.
Eggebrecht  H., Kuhl  H., Kaiser  G.M.; Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J. 27 2006:613-620.
Krueger  J.J., Ewert  P., Yilmaz  S.; Magnetic resonance imaging-guided balloon angioplasty of coarctation of the aorta: a pilot study. Circulation. 113 2006:1093-1100.
Raval  A.N., Karmarkar  P.V., Guttman  M.A.; Real-time magnetic resonance imaging-guided endovascular recanalization of chronic total arterial occlusion in a swine model. Circulation. 113 2006:1101-1107.
Anderson  K.J., Leung  G., Dick  A.J., Wright  G.A.; Forward-looking intravascular orthogonal-solenoid coil for imaging and guidance in occlusive arterial disease. Magn Reson Med. 60 2008:489-495.
Kee  S.T., Ganguly  A., Daniel  B.L.; MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system. J Vasc Interv Radiol. 16 2005:227-234.
Arepally  A., Karmarkar  P.V., Weiss  C., Atalar  E.; Percutaneous MR imaging-guided transvascular access of mesenteric venous system: study in swine model. Radiology. 238 2006:113-118.
Raval  A.N., Karmarkar  P.V., Guttman  M.A.; Real-time MRI guided atrial septal puncture and balloon septostomy in swine. Catheter Cardiovasc Interv. 67 2006:637-643.
Elagha  A.A., Kocaturk  O., Guttman  M.A.; Real-time MR imaging-guided laser atrial septal puncture in swine. J Vasc Interv Radiol. 19 2008:1347-1353.
Buecker  A., Spuentrup  E., Grabitz  R.; Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale. Circulation. 106 2002:511-515.
Rickers  C., Jerosch-Herold  M., Hu  X.; Magnetic resonance image-guided transcatheter closure of atrial septal defects. Circulation. 107 2003:132-138.
Schalla  S., Saeed  M., Higgins  C.B., Weber  O., Martin  A., Moore  P.; Balloon sizing and transcatheter closure of acute atrial septal defects guided by magnetic resonance fluoroscopy: assessment and validation in a large animal model. J Magn Reson Imaging. 21 2005:204-211.
Kuehne  T., Yilmaz  S., Meinus  C.; Magnetic resonance imaging-guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 44 2004:2247-2249.
McVeigh  E.R., Guttman  M.A., Lederman  R.J.; Real-time interactive MRI-guided cardiac surgery: aortic valve replacement using a direct apical approach. Magn Reson Med. 56 2006:958-964.
Bracken  J.A., Komljenovic  P., Lillaney  P.V., Fahrig  R., Rowlands  J.A.; Closed-bore XMR (CBXMR) systems for aortic valve replacement: X-ray tube imaging performance. Med Phys. 36 2009:1086-1097.
Lederman  R.J., Guttman  M.A., Peters  D.C.; Catheter-based endomyocardial injection with real-time magnetic resonance imaging. Circulation. 105 2002:1282-1284.
Corti  R., Badimon  J., Mizsei  G.; Real time magnetic resonance guided endomyocardial local delivery. Heart. 91 2005:348-353.
Barbash  I.M., Leor  J., Feinberg  M.S.; Interventional magnetic resonance imaging for guiding gene and cell transfer in the heart. Heart. 90 2004:87-91.
Dick  A.J., Guttman  M.A., Raman  V.K.; Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine. Circulation. 108 2003:2899-2904.
Saeed  M., Martin  A., Jacquier  A.; Permanent coronary artery occlusion: cardiovascular MR imaging is platform for percutaneous transendocardial delivery and assessment of gene therapy in canine model. Radiology. 249 2008:560-571.
Kraitchman  D.L., Heldman  A.W., Atalar  E.; In vivo magnetic resonance imaging of mesenchymal stem cells in myocardial infarction. Circulation. 107 2003:2290-2293.
Andreassi  M.G., Ait-Ali  L., Botto  N., Manfredi  S., Mottola  G., Picano  E.; Cardiac catheterization and long-term chromosomal damage in children with congenital heart disease. Eur Heart J. 27 2006:2703-2708.
Vano  E., Gonzalez  L., Fernandez  J.M., Haskal  Z.J.; Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology. 248 2008:945-953.
Goldstein  J.A., Balter  S., Cowley  M., Hodgson  J., Klein  L.W.; Occupational hazards of interventional cardiologists: prevalence of orthopedic health problems in contemporary practice. Catheter Cardiovasc Interv. 63 2004:407-411.
Razavi  R., Hill  D.L., Keevil  S.F.; Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. Lancet. 362 2003:1877-1882.
Muthurangu  V., Taylor  A., Andriantsimiavona  R.; Novel method of quantifying pulmonary vascular resistance by use of simultaneous invasive pressure monitoring and phase-contrast magnetic resonance flow. Circulation. 110 2004:826-834.
Kuehne  T., Yilmaz  S., Schulze-Neick  I.; Magnetic resonance imaging guided catheterisation for assessment of pulmonary vascular resistance: in vivo validation and clinical application in patients with pulmonary hypertension. Heart. 91 2005:1064-1069.
Kuehne  T., Saeed  M., Higgins  C.B.; Endovascular stents in pulmonary valve and artery in swine: feasibility study of MR imaging-guided deployment and postinterventional assessment. Radiology. 226 2003:475-481.
Khambadkone  S., Coats  L., Taylor  A.; Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation. 112 2005:1189-1197.
Dickfeld  T., Calkins  H., Zviman  M.; Anatomic stereotactic catheter ablation on three-dimensional magnetic resonance images in real time. Circulation. 108 2003:2407-2413.
Dickfeld  T., Kato  R., Zviman  M.; Characterization of radiofrequency ablation lesions with gadolinium-enhanced cardiovascular magnetic resonance imaging. J Am Coll Cardiol. 47 2006:370-378.
McGann  C.J., Kholmovski  E.G., Oakes  R.S.; New magnetic resonance imaging-based method for defining the extent of left atrial wall injury after the ablation of atrial fibrillation. J Am Coll Cardiol. 52 2008:1263-1271.
Susil  R.C., Yeung  C.J., Halperin  H.R., Lardo  A.C., Atalar  E.; Multifunctional interventional devices for MRI: a combined electrophysiology/MRI catheter. Magn Reson Med. 47 2002:594-600.
Dukkipati  S.R., Mallozzi  R., Schmidt  E.J.; Electroanatomic mapping of the left ventricle in a porcine model of chronic myocardial infarction with magnetic resonance-based catheter tracking. Circulation. 118 2008:853-862.
Krueger  S., Lips  O., David  B.; Towards MR-guided EP interventions using an RF-safe concept. J Cardiovasc Magn Reson. 11 (Suppl) 2009:O84
Nazarian  S., Kolandaivelu  A., Zviman  M.M.; Feasibility of real-time magnetic resonance imaging for catheter guidance in electrophysiology studies. Circulation. 118 2008:223-229.
Gutierrez  L.F., Silva  R., Ozturk  C.; Technology preview: X-ray fused with magnetic resonance during invasive cardiovascular procedures. Catheter Cardiovasc Interv. 70 2007:773-782.
de Silva  R., Gutierrez  L.F., Raval  A.N., McVeigh  E.R., Ozturk  C., Lederman  R.J.; X-ray fused with magnetic resonance imaging (XFM) to target endomyocardial injections: validation in a swine model of myocardial infarction. Circulation. 114 2006:2342-2350.
Ratnayaka  K., Raman  V.K., Faranesh  A.Z.; Antegrade percutaneous closure of membranous ventricular septal defect using X-ray fused with magnetic resonance imaging. J Am Coll Cardiol Intv. 2 2009:224-230.
Rhode  K.S., Hill  D.L., Edwards  P.J.; Registration and tracking to integrate X-ray and MR images in an XMR facility. IEEE Trans Med Imaging. 22 2003:1369-1378.
Rhode  K.S., Sermesant  M., Brogan  D.; A system for real-time XMR-guided cardiovascular intervention. IEEE Trans Med Imaging. 24 2005:1428-1440.
King  A.P., Boubertakh  R., Rhode  K.S.; A subject-specific technique for respiratory motion correction in image-guided cardiac catheterisation procedures. Med Image Anal. 13 2009:419-431.
Tronnier  V.M., Wirtz  C.R., Knauth  M.; Intraoperative diagnostic and interventional magnetic resonance imaging in neurosurgery. Neurosurgery. 40 1997:891-902.
Martin  A.J., Larson  P.S., Ostrem  J.L.; Placement of deep brain stimulator electrodes using real-time high-field interventional magnetic resonance imaging. Magn Reson Med. 54 2005:1107-1114.
Weiss  C.R., Nour  S.G., Lewin  J.S.; MR-guided biopsy: a review of current techniques and applications. J Magn Reson Imaging. 27 2008:311-325.
Rieke  V., Butts Pauly  K.; MR thermometry. J Magn Reson Imaging. 27 2008:376-390.
Smith  K.A., Carrino  J.; MRI-guided interventions of the musculoskeletal system. J Magn Reson Imaging. 27 2008:339-346.
Cleary  K., Melzer  A., Watson  V., Kronreif  G., Stoianovici  D.; Interventional robotic systems: applications and technology state-of-the-art. Minim Invasive Ther Allied Technol. 15 2006:101-113.
Hoult  D.I., Saunders  J.K., Sutherland  G.R.; The engineering of an interventional MRI with a movable 1.5 Tesla magnet. J Magn Reson Imaging. 13 2001:78-86.
Griswold  M.A., Jakob  P.M., Heidemann  R.M.; Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 47 2002:1202-1210.
Kellman  P., Epstein  F.H., McVeigh  E.R.; Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 45 2001:846-852.
Pruessmann  K.P., Weiger  M., Scheidegger  M.B., Boesiger  P.; SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 42 1999:952-962.
Guttman  M.A., Ozturk  C., Raval  A.N.; Interventional cardiovascular procedures guided by real-time MR imaging: an interactive interface using multiple slices, adaptive projection modes and live 3D renderings. J Magn Reson Imaging. 26 2007:1429-1435.
Santos  J.M., Wright  G.A., Pauly  J.M.; Flexible real-time magnetic resonance imaging framework. Conf Proc IEEE Eng Med Biol Soc. 2 2004:1048-1051.
Yutzy  S.R., Duerk  J.L.; Pulse sequences and system interfaces for interventional and real-time MRI. J Magn Reson Imaging. 27 2008:267-275.
Schirra  C.O., Weiss  S., Krueger  S.; Toward true 3D visualization of active catheters using compressed sensing. Magn Reson Med. 62 2009:341-347.
George  A., Derbyshire  J., Saybasili  H.; Fast, robust 3D visualization and automatic slice repositioning (“snap-to”) for MR-guided interventions using active device “profiling.”. Paper presented at: ISMRM 17th Annual Scientific Meeting April 18 to 24 2009 Honolulu, HI
Elgort  D.R., Wong  E.Y., Hillenbrand  C.M., Wacker  F.K., Lewin  J.S., Duerk  J.L.; Real-time catheter tracking and adaptive imaging. J Magn Reson Imaging. 18 2003:621-626.
Patil  S., Bieri  O., Jhooti  P., Scheffler  K.; Automatic slice positioning (ASP) for passive real-time tracking of interventional devices using projection-reconstruction imaging with echo-dephasing (PRIDE). Magn Reson Med. 62 2009:935-942.
Rubin  D.L., Ratner  A.V., Young  S.W.; Magnetic susceptibility effects and their application in the development of new ferromagnetic catheters for magnetic resonance imaging. Invest Radiol. 25 1990:1325-1332.
Bakker  C.J., Hoogeveen  R.M., Weber  J., van Vaals  J.J., Viergever  M.A., Mali  W.P.; Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions. Magn Reson Med. 36 1996:816-820.
Omary  R.A., Unal  O., Koscielski  D.S.; Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters. J Vasc Interv Radiol. 11 2000:1079-1085.
Miquel  M.E., Hegde  S., Muthurangu  V.; Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patients. Magn Reson Med. 51 2004:988-995.
Kozerke  S., Hegde  S., Schaeffter  T., Lamerichs  R., Razavi  R., Hill  D.L.; Catheter tracking and visualization using 19F nuclear magnetic resonance. Magn Reson Med. 52 2004:693-697.
Magnusson  P., Johansson  E., Mansson  S.; Passive catheter tracking during interventional MRI using hyperpolarized 13C. Magn Reson Med. 57 2007:1140-1147.
Dharmakumar  R., Koktzoglou  I., Tang  R., Harris  K.R., Beohar  N., Li  D.; Off-resonance positive contrast imaging of a passive endomyocardial catheter in swine. Phys Med Biol. 53 2008:N249-N257.
Edelman  R.R., Storey  P., Dunkle  E.; Gadolinium-enhanced off-resonance contrast angiography. Magn Reson Med. 57 2007:475-484.
Krueger  S., Schmitz  S., Weiss  S.; An MR guidewire based on micropultruded fiber-reinforced material. Magn Reson Med. 60 2008:1190-1196.
Mekle  R., Hofmann  E., Scheffler  K., Bilecen  D.; A polymer-based MR-compatible guidewire: a study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). J Magn Reson Imaging. 23 2006:145-155.
Dumoulin  C.L., Souza  S.P., Darrow  R.D.; Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med. 29 1993:411-415.
Ocali  O., Atalar  E.; Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med. 37 1997:112-118.
Kocaturk  O., Kim  A.H., Saikus  C.E.; Active two-channel 0.035'' guidewire for interventional cardiovascular MRI. J Magn Reson Imaging. 30 2009:461-465.
Kuehne  T., Fahrig  R., Butts  K.; Pair of resonant fiducial markers for localization of endovascular catheters at all catheter orientations. J Magn Reson Imaging. 17 2003:620-624.
Quick  H.H., Zenge  M.O., Kuehl  H.; Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med. 53 2005:446-455.
Eggers  H., Weiss  S., Boernert  P., Boesiger  P.; Image-based tracking of optically detunable parallel resonant circuits. Magn Reson Med. 49 2003:1163-1174.
Wong  E.Y., Zhang  Q., Duerk  J.L., Lewin  J.S., Wendt  M.; An optical system for wireless detuning of parallel resonant circuits. J Magn Reson Imaging. 12 2000:632-638.
Celik  H., Uluturk  A., Tali  T., Atalar  E.; A catheter tracking method using reverse polarization for MR-guided interventions. Magn Reson Med. 58 2007:1224-1231.
Pictet  J., Meuli  R., Wicky  S., van der Klink  J.J.; Radiofrequency heating effects around resonant lengths of wire in MRI. Phys Med Biol. 47 2002:2973-2985.
Ladd  M.E., Quick  H.H.; Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magn Reson Med. 43 2000:615-619.
Weiss  S., Vernickel  P., Schaeffter  T., Schulz  V., Gleich  B.; Transmission line for improved RF safety of interventional devices. Magn Reson Med. 54 2005:182-189.
Fandrey  S., Weiss  S., Muller  J.; Development of an active intravascular MR device with an optical transmission system. IEEE Trans Med Imaging. 27 2008:1723-1727.

Figures

Grahic Jump Location
Figure 1

CMR-Guided Transapical Aortic Valve Replacement

(A) Bioprosthesis mounted on a platinum iridium stent with a stainless-steel marker welded on the side of the stent between the commissures. (B) The marker is visible as a dark signal in the cardiovascular magnetic resonance (CMR) and indicates the orientation of the prosthesis. Short-axis view and long-axis view of the implanted prosthesis in a pig under real-time CMR are shown in C and D, respectively. Blue dots are digital markers indicating the coronary ostia whereas the yellow dot shows aortic annulus location. (E) Three-dimensional rendering snapshots show multiple image planes displayed at their relative 3-dimensional position. Images courtesy of Ming Li, PhD, and Keith A. Horvath, MD, Cardiothoracic Surgery Research Program, National Heart, Lung, and Blood Institute.

Grahic Jump Location
Figure 2

Electrophysiological Cardiac Mapping Under Real-Time CMR Guidance

(Top) Roadmap-based real-time 3-dimensional visualization of the catheter position during recording (red dot) on the magnetic resonance-electrophysiological workstation. The yellow dots in the 3-dimensional rendering of the heart indicate previous mapping positions. (Bottom) In-bore electrophysiological recordings at 2 selected positions showing an atrial signal (left) and a ventricular signal (right). Graphic courtesy of Steffen Weiss, PhD, Philips Healthcare.

Grahic Jump Location
Figure 3

X-Ray Fused With MRI

Representative X-ray fused with magnetic resonance imaging (MRI) display with contours from cardiovascular magnetic resonance (CMR) overlaid on live X-ray. Aorta (red), right ventricle (yellow), left endocardial (blue), and epicardial (green) surfaces from CMR shown in 3 dimensions with aortic annulus and ventricular septal defect locations outlined in blue and red, respectively. Multimodality external fiducial markers also are seen numbered.

Grahic Jump Location
Figure 4

Interventional CMR Suite

An interventional cardiovascular magnetic resonance (CMR) suite showing adjacent and interoperable magnetic resonance imaging scanner and X-ray angiography labs. A docking gurney allows rapid intermodality transfer. Courtesy of Alexander J. Dick, MD, Sunnybrook Health Sciences Centre, Toronto.

Grahic Jump Location
Figure 5

Real-Time Cardiac MR

Multiplanar real-time cardiac imaging at 4 to 5 frames/s (A) with pre-acquired static cardiac magnetic resonance (MR) image (B) and pulmonary vessel 3-dimensional MR angiography (C) roadmaps.

Grahic Jump Location
Figure 6

Real-Time MRI User Interface Showing MR-Guided Delivery and Biopsy of Encapsulated Human Islet Cells in a Pig

Example of a real-time magnetic resonance imaging (MRI) user interface from Siemens. Similar interfaces are available for Philips and GE systems. Features include scan plane manipulation, independent device channel coloring, 3-dimensional viewing window, static image roadmaps, and quick access to select slice orientations. Figure courtesy of Christine H. Lorenz, PhD, Siemens Corporate Research, Inc., and Aravind Arepally, MD, Johns Hopkins Medical Institutions, Baltimore, Maryland.

Grahic Jump Location
Figure 7

Interventional CMR Devices

Three representative approaches to catheters designed to be tracked using cardiovascular magnetic resonance (CMR). (A) Passive nonbraided balloon catheter filled with CO2 in a right atrium. (B) Active tracking catheter in the right heart. Image courtesy of Michael Bock, PhD, German Cancer Research Center (DKFZ), Heidelberg, Germany. (C) Two-channel active guidewire in the aorta approaching the left heart.

Grahic Jump Location
Figure 8

Iliac Artery Navigation Using a Passive Guidewire

(Top) Polymer guidewire with a Nitinol tip. (Bottom) Magnetic resonance-guided probing of the contralateral iliac artery with the guidewire (arrows) and a catheter. Susceptibility artifacts in cardiovascular magnetic resonance (CMR) images arise from the iron doped guidewire, which is also visible under fluoroscopy. These artifacts partially obscure the target tissue. Images courtesy of Gabriele A. Krombach, MD, Department of Diagnostic Radiology, University Hospital Aachen, Aachen, Germany.

Tables

Table Grahic Jump Location
Table 1Comparison of Imaging Modalities for Real-Time Procedural Guidance
Table Footer NoteNote the frame rate and spatial resolution given for CMR are arbitrary and represent a typical compromise between spatial and temporal resolution at 1.5-T using parallel imaging with an acceleration factor of 2.

Interactive Graphics

Video

References

Kantor  H.L., Briggs  R.W., Balaban  R.S.; In vivo 31P nuclear magnetic resonance measurements in canine heart using a catheter-coil. Circ Res. 55 1984:261-266.
Hurst  G.C., Hua  J., Duerk  J.L., Cohen  A.M.; Intravascular (catheter) NMR receiver probe: preliminary design analysis and application to canine iliofemoral imaging. Magn Reson Med. 24 1992:343-357.
Martin  A.J., Plewes  D.B., Henkelman  R.M.; MR imaging of blood vessels with an intravascular coil. J Magn Reson Imaging. 2 1992:421-429.
Nordbeck  P., Bauer  W.R., Fidler  F.; Feasibility of real-time MRI with a novel carbon catheter for interventional electrophysiology. Circ Arrhythmia Electrophysiol. 2 2009:258-267.
Raval  A.N., Telep  J.D., Guttman  M.A.; Real-time magnetic resonance imaging-guided stenting of aortic coarctation with commercially available catheter devices in swine. Circulation. 112 2005:699-706.
Omary  R.A., Green  J.D., Schirf  B.E., Li  Y., Finn  J.P., Li  D.; Real-time magnetic resonance imaging-guided coronary catheterization in swine. Circulation. 107 2003:2656-2659.
Serfaty  J.M., Yang  X., Foo  T.K., Kumar  A., Derbyshire  A., Atalar  E.; MRI-guided coronary catheterization and PTCA: a feasibility study on a dog model. Magn Reson Med. 49 2003:258-263.
Spuentrup  E., Ruebben  A., Schaeffter  T., Manning  W.J., Gunther  R.W., Buecker  A.; Magnetic resonance-guided coronary artery stent placement in a swine model. Circulation. 105 2002:874-879.
Manke  C., Nitz  W.R., Djavidani  B.; MR imaging-guided stent placement in iliac arterial stenoses: a feasibility study. Radiology. 219 2001:527-534.
Elgort  D.R., Hillenbrand  C.M., Zhang  S.; Image-guided and -monitored renal artery stenting using only MRI. J Magn Reson Imaging. 23 2006:619-627.
Raman  V.K., Karmarkar  P.V., Guttman  M.A.; Real-time magnetic resonance-guided endovascular repair of experimental abdominal aortic aneurysm in swine. J Am Coll Cardiol. 45 2005:2069-2077.
Eggebrecht  H., Kuhl  H., Kaiser  G.M.; Feasibility of real-time magnetic resonance-guided stent-graft placement in a swine model of descending aortic dissection. Eur Heart J. 27 2006:613-620.
Krueger  J.J., Ewert  P., Yilmaz  S.; Magnetic resonance imaging-guided balloon angioplasty of coarctation of the aorta: a pilot study. Circulation. 113 2006:1093-1100.
Raval  A.N., Karmarkar  P.V., Guttman  M.A.; Real-time magnetic resonance imaging-guided endovascular recanalization of chronic total arterial occlusion in a swine model. Circulation. 113 2006:1101-1107.
Anderson  K.J., Leung  G., Dick  A.J., Wright  G.A.; Forward-looking intravascular orthogonal-solenoid coil for imaging and guidance in occlusive arterial disease. Magn Reson Med. 60 2008:489-495.
Kee  S.T., Ganguly  A., Daniel  B.L.; MR-guided transjugular intrahepatic portosystemic shunt creation with use of a hybrid radiography/MR system. J Vasc Interv Radiol. 16 2005:227-234.
Arepally  A., Karmarkar  P.V., Weiss  C., Atalar  E.; Percutaneous MR imaging-guided transvascular access of mesenteric venous system: study in swine model. Radiology. 238 2006:113-118.
Raval  A.N., Karmarkar  P.V., Guttman  M.A.; Real-time MRI guided atrial septal puncture and balloon septostomy in swine. Catheter Cardiovasc Interv. 67 2006:637-643.
Elagha  A.A., Kocaturk  O., Guttman  M.A.; Real-time MR imaging-guided laser atrial septal puncture in swine. J Vasc Interv Radiol. 19 2008:1347-1353.
Buecker  A., Spuentrup  E., Grabitz  R.; Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale. Circulation. 106 2002:511-515.
Rickers  C., Jerosch-Herold  M., Hu  X.; Magnetic resonance image-guided transcatheter closure of atrial septal defects. Circulation. 107 2003:132-138.
Schalla  S., Saeed  M., Higgins  C.B., Weber  O., Martin  A., Moore  P.; Balloon sizing and transcatheter closure of acute atrial septal defects guided by magnetic resonance fluoroscopy: assessment and validation in a large animal model. J Magn Reson Imaging. 21 2005:204-211.
Kuehne  T., Yilmaz  S., Meinus  C.; Magnetic resonance imaging-guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 44 2004:2247-2249.
McVeigh  E.R., Guttman  M.A., Lederman  R.J.; Real-time interactive MRI-guided cardiac surgery: aortic valve replacement using a direct apical approach. Magn Reson Med. 56 2006:958-964.
Bracken  J.A., Komljenovic  P., Lillaney  P.V., Fahrig  R., Rowlands  J.A.; Closed-bore XMR (CBXMR) systems for aortic valve replacement: X-ray tube imaging performance. Med Phys. 36 2009:1086-1097.
Lederman  R.J., Guttman  M.A., Peters  D.C.; Catheter-based endomyocardial injection with real-time magnetic resonance imaging. Circulation. 105 2002:1282-1284.
Corti  R., Badimon  J., Mizsei  G.; Real time magnetic resonance guided endomyocardial local delivery. Heart. 91 2005:348-353.
Barbash  I.M., Leor  J., Feinberg  M.S.; Interventional magnetic resonance imaging for guiding gene and cell transfer in the heart. Heart. 90 2004:87-91.
Dick  A.J., Guttman  M.A., Raman  V.K.; Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine. Circulation. 108 2003:2899-2904.
Saeed  M., Martin  A., Jacquier  A.; Permanent coronary artery occlusion: cardiovascular MR imaging is platform for percutaneous transendocardial delivery and assessment of gene therapy in canine model. Radiology. 249 2008:560-571.
Kraitchman  D.L., Heldman  A.W., Atalar  E.; In vivo magnetic resonance imaging of mesenchymal stem cells in myocardial infarction. Circulation. 107 2003:2290-2293.
Andreassi  M.G., Ait-Ali  L., Botto  N., Manfredi  S., Mottola  G., Picano  E.; Cardiac catheterization and long-term chromosomal damage in children with congenital heart disease. Eur Heart J. 27 2006:2703-2708.
Vano  E., Gonzalez  L., Fernandez  J.M., Haskal  Z.J.; Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology. 248 2008:945-953.
Goldstein  J.A., Balter  S., Cowley  M., Hodgson  J., Klein  L.W.; Occupational hazards of interventional cardiologists: prevalence of orthopedic health problems in contemporary practice. Catheter Cardiovasc Interv. 63 2004:407-411.
Razavi  R., Hill  D.L., Keevil  S.F.; Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. Lancet. 362 2003:1877-1882.
Muthurangu  V., Taylor  A., Andriantsimiavona  R.; Novel method of quantifying pulmonary vascular resistance by use of simultaneous invasive pressure monitoring and phase-contrast magnetic resonance flow. Circulation. 110 2004:826-834.
Kuehne  T., Yilmaz  S., Schulze-Neick  I.; Magnetic resonance imaging guided catheterisation for assessment of pulmonary vascular resistance: in vivo validation and clinical application in patients with pulmonary hypertension. Heart. 91 2005:1064-1069.
Kuehne  T., Saeed  M., Higgins  C.B.; Endovascular stents in pulmonary valve and artery in swine: feasibility study of MR imaging-guided deployment and postinterventional assessment. Radiology. 226 2003:475-481.
Khambadkone  S., Coats  L., Taylor  A.; Percutaneous pulmonary valve implantation in humans: results in 59 consecutive patients. Circulation. 112 2005:1189-1197.
Dickfeld  T., Calkins  H., Zviman  M.; Anatomic stereotactic catheter ablation on three-dimensional magnetic resonance images in real time. Circulation. 108 2003:2407-2413.
Dickfeld  T., Kato  R., Zviman  M.; Characterization of radiofrequency ablation lesions with gadolinium-enhanced cardiovascular magnetic resonance imaging. J Am Coll Cardiol. 47 2006:370-378.
McGann  C.J., Kholmovski  E.G., Oakes  R.S.; New magnetic resonance imaging-based method for defining the extent of left atrial wall injury after the ablation of atrial fibrillation. J Am Coll Cardiol. 52 2008:1263-1271.
Susil  R.C., Yeung  C.J., Halperin  H.R., Lardo  A.C., Atalar  E.; Multifunctional interventional devices for MRI: a combined electrophysiology/MRI catheter. Magn Reson Med. 47 2002:594-600.
Dukkipati  S.R., Mallozzi  R., Schmidt  E.J.; Electroanatomic mapping of the left ventricle in a porcine model of chronic myocardial infarction with magnetic resonance-based catheter tracking. Circulation. 118 2008:853-862.
Krueger  S., Lips  O., David  B.; Towards MR-guided EP interventions using an RF-safe concept. J Cardiovasc Magn Reson. 11 (Suppl) 2009:O84
Nazarian  S., Kolandaivelu  A., Zviman  M.M.; Feasibility of real-time magnetic resonance imaging for catheter guidance in electrophysiology studies. Circulation. 118 2008:223-229.
Gutierrez  L.F., Silva  R., Ozturk  C.; Technology preview: X-ray fused with magnetic resonance during invasive cardiovascular procedures. Catheter Cardiovasc Interv. 70 2007:773-782.
de Silva  R., Gutierrez  L.F., Raval  A.N., McVeigh  E.R., Ozturk  C., Lederman  R.J.; X-ray fused with magnetic resonance imaging (XFM) to target endomyocardial injections: validation in a swine model of myocardial infarction. Circulation. 114 2006:2342-2350.
Ratnayaka  K., Raman  V.K., Faranesh  A.Z.; Antegrade percutaneous closure of membranous ventricular septal defect using X-ray fused with magnetic resonance imaging. J Am Coll Cardiol Intv. 2 2009:224-230.
Rhode  K.S., Hill  D.L., Edwards  P.J.; Registration and tracking to integrate X-ray and MR images in an XMR facility. IEEE Trans Med Imaging. 22 2003:1369-1378.
Rhode  K.S., Sermesant  M., Brogan  D.; A system for real-time XMR-guided cardiovascular intervention. IEEE Trans Med Imaging. 24 2005:1428-1440.
King  A.P., Boubertakh  R., Rhode  K.S.; A subject-specific technique for respiratory motion correction in image-guided cardiac catheterisation procedures. Med Image Anal. 13 2009:419-431.
Tronnier  V.M., Wirtz  C.R., Knauth  M.; Intraoperative diagnostic and interventional magnetic resonance imaging in neurosurgery. Neurosurgery. 40 1997:891-902.
Martin  A.J., Larson  P.S., Ostrem  J.L.; Placement of deep brain stimulator electrodes using real-time high-field interventional magnetic resonance imaging. Magn Reson Med. 54 2005:1107-1114.
Weiss  C.R., Nour  S.G., Lewin  J.S.; MR-guided biopsy: a review of current techniques and applications. J Magn Reson Imaging. 27 2008:311-325.
Rieke  V., Butts Pauly  K.; MR thermometry. J Magn Reson Imaging. 27 2008:376-390.
Smith  K.A., Carrino  J.; MRI-guided interventions of the musculoskeletal system. J Magn Reson Imaging. 27 2008:339-346.
Cleary  K., Melzer  A., Watson  V., Kronreif  G., Stoianovici  D.; Interventional robotic systems: applications and technology state-of-the-art. Minim Invasive Ther Allied Technol. 15 2006:101-113.
Hoult  D.I., Saunders  J.K., Sutherland  G.R.; The engineering of an interventional MRI with a movable 1.5 Tesla magnet. J Magn Reson Imaging. 13 2001:78-86.
Griswold  M.A., Jakob  P.M., Heidemann  R.M.; Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 47 2002:1202-1210.
Kellman  P., Epstein  F.H., McVeigh  E.R.; Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 45 2001:846-852.
Pruessmann  K.P., Weiger  M., Scheidegger  M.B., Boesiger  P.; SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 42 1999:952-962.
Guttman  M.A., Ozturk  C., Raval  A.N.; Interventional cardiovascular procedures guided by real-time MR imaging: an interactive interface using multiple slices, adaptive projection modes and live 3D renderings. J Magn Reson Imaging. 26 2007:1429-1435.
Santos  J.M., Wright  G.A., Pauly  J.M.; Flexible real-time magnetic resonance imaging framework. Conf Proc IEEE Eng Med Biol Soc. 2 2004:1048-1051.
Yutzy  S.R., Duerk  J.L.; Pulse sequences and system interfaces for interventional and real-time MRI. J Magn Reson Imaging. 27 2008:267-275.
Schirra  C.O., Weiss  S., Krueger  S.; Toward true 3D visualization of active catheters using compressed sensing. Magn Reson Med. 62 2009:341-347.
George  A., Derbyshire  J., Saybasili  H.; Fast, robust 3D visualization and automatic slice repositioning (“snap-to”) for MR-guided interventions using active device “profiling.”. Paper presented at: ISMRM 17th Annual Scientific Meeting April 18 to 24 2009 Honolulu, HI
Elgort  D.R., Wong  E.Y., Hillenbrand  C.M., Wacker  F.K., Lewin  J.S., Duerk  J.L.; Real-time catheter tracking and adaptive imaging. J Magn Reson Imaging. 18 2003:621-626.
Patil  S., Bieri  O., Jhooti  P., Scheffler  K.; Automatic slice positioning (ASP) for passive real-time tracking of interventional devices using projection-reconstruction imaging with echo-dephasing (PRIDE). Magn Reson Med. 62 2009:935-942.
Rubin  D.L., Ratner  A.V., Young  S.W.; Magnetic susceptibility effects and their application in the development of new ferromagnetic catheters for magnetic resonance imaging. Invest Radiol. 25 1990:1325-1332.
Bakker  C.J., Hoogeveen  R.M., Weber  J., van Vaals  J.J., Viergever  M.A., Mali  W.P.; Visualization of dedicated catheters using fast scanning techniques with potential for MR-guided vascular interventions. Magn Reson Med. 36 1996:816-820.
Omary  R.A., Unal  O., Koscielski  D.S.; Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters. J Vasc Interv Radiol. 11 2000:1079-1085.
Miquel  M.E., Hegde  S., Muthurangu  V.; Visualization and tracking of an inflatable balloon catheter using SSFP in a flow phantom and in the heart and great vessels of patients. Magn Reson Med. 51 2004:988-995.
Kozerke  S., Hegde  S., Schaeffter  T., Lamerichs  R., Razavi  R., Hill  D.L.; Catheter tracking and visualization using 19F nuclear magnetic resonance. Magn Reson Med. 52 2004:693-697.
Magnusson  P., Johansson  E., Mansson  S.; Passive catheter tracking during interventional MRI using hyperpolarized 13C. Magn Reson Med. 57 2007:1140-1147.
Dharmakumar  R., Koktzoglou  I., Tang  R., Harris  K.R., Beohar  N., Li  D.; Off-resonance positive contrast imaging of a passive endomyocardial catheter in swine. Phys Med Biol. 53 2008:N249-N257.
Edelman  R.R., Storey  P., Dunkle  E.; Gadolinium-enhanced off-resonance contrast angiography. Magn Reson Med. 57 2007:475-484.
Krueger  S., Schmitz  S., Weiss  S.; An MR guidewire based on micropultruded fiber-reinforced material. Magn Reson Med. 60 2008:1190-1196.
Mekle  R., Hofmann  E., Scheffler  K., Bilecen  D.; A polymer-based MR-compatible guidewire: a study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). J Magn Reson Imaging. 23 2006:145-155.
Dumoulin  C.L., Souza  S.P., Darrow  R.D.; Real-time position monitoring of invasive devices using magnetic resonance. Magn Reson Med. 29 1993:411-415.
Ocali  O., Atalar  E.; Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med. 37 1997:112-118.
Kocaturk  O., Kim  A.H., Saikus  C.E.; Active two-channel 0.035'' guidewire for interventional cardiovascular MRI. J Magn Reson Imaging. 30 2009:461-465.
Kuehne  T., Fahrig  R., Butts  K.; Pair of resonant fiducial markers for localization of endovascular catheters at all catheter orientations. J Magn Reson Imaging. 17 2003:620-624.
Quick  H.H., Zenge  M.O., Kuehl  H.; Interventional magnetic resonance angiography with no strings attached: wireless active catheter visualization. Magn Reson Med. 53 2005:446-455.
Eggers  H., Weiss  S., Boernert  P., Boesiger  P.; Image-based tracking of optically detunable parallel resonant circuits. Magn Reson Med. 49 2003:1163-1174.
Wong  E.Y., Zhang  Q., Duerk  J.L., Lewin  J.S., Wendt  M.; An optical system for wireless detuning of parallel resonant circuits. J Magn Reson Imaging. 12 2000:632-638.
Celik  H., Uluturk  A., Tali  T., Atalar  E.; A catheter tracking method using reverse polarization for MR-guided interventions. Magn Reson Med. 58 2007:1224-1231.
Pictet  J., Meuli  R., Wicky  S., van der Klink  J.J.; Radiofrequency heating effects around resonant lengths of wire in MRI. Phys Med Biol. 47 2002:2973-2985.
Ladd  M.E., Quick  H.H.; Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magn Reson Med. 43 2000:615-619.
Weiss  S., Vernickel  P., Schaeffter  T., Schulz  V., Gleich  B.; Transmission line for improved RF safety of interventional devices. Magn Reson Med. 54 2005:182-189.
Fandrey  S., Weiss  S., Muller  J.; Development of an active intravascular MR device with an optical transmission system. IEEE Trans Med Imaging. 27 2008:1723-1727.

Correspondence

Latest JACC CME

Continuing Medical Education through JACC is a convenient way to fulfill your CME requirements while learning important information about the latest advances in cardiovascular medicine.

April 2013- JACC CME Activity
Repeat Revascularization and Outcome

March 2013- JACC CME Activity
Extreme Lipoprotein(a) Levels and Improved Cardiovascular Risk Prediction

Feb 2013- JACC CME Activity
Results from the BARI 2D Trial

Jan 2013- JACC CME Activity
Prognosis Among Healthy Individuals Discharged With a Primary Diagnosis of Syncope

Dec 2012- JACC CME Activity
Incidence of Heart Failure or Cardiomyopathy After Adjuvant Trastuzumab Therapy for Breast Cancer

Nov 2012- JACC CME Activity
A Collaborative Analysis of Individual Patient Data From 10 Randomized Trials

Oct 2012- JACC CME Activity
Radiofrequency Ablation of Premature Ventricular Ectopy Improves the Efficacy of Cardiac Resynchronization Therapy in Nonresponders

Sept 2012- JACC CME Activity
Exercise and Pharmacological Treatment of Depressive Symptoms in Patients With Coronary Heart Disease

Aug 2012- JACC CME Activity
Reduction in Life-Threatening Ventricular Tachyarrhythmias in Statin-Treated Patients With Nonischemic Cardiomyopathy Enrolled in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy)

July 2012- JACC CME Activity
Relationship of Beta-Blocker Dose With Outcomes in Ambulatory Heart Failure Patients With Systolic Dysfunction

For previous CME quizzes, please follow this link to CardioSource Lifelong Learning and MOC.

 

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s “Cited By” API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment
Submit a Comment

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics